Skip to content
Recall Observatory FDA recall evidence

Device product

Ingenuity Core, (Code 728321) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Z-0390-2018

November 10, 2017

Class II

Product summary

Firm
Philips Medical Systems (Cleveland) Inc
Event
Event 78852
Status
Terminated
Classification
Class II
Quantity
1
Official record key
device-enforcement:Z-0390-2018

Official wording

Reason: Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

Code information: Code 728321, SN # 310094

Distribution pattern: Worldwide Distribution -- US, Argentina, China, Republic of Korea, Russian Federation, Slovakia, Turkey, and United Kingdom. 4 in US, (1 military), 14 foreign accounts

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.