Skip to content
Recall Observatory FDA recall evidence

Device product

SORIN GROUP, SMART PERFUSION PACK, SMART CIRCULATORY ASSIST PACK, REF 627255003, STERILE EO, Rx Only, 1 EA

Z-2784-2017

June 29, 2017

Class II

Product summary

Firm
Sorin Group USA, Inc.
Event
Event 77810
Status
Terminated
Classification
Class II
Quantity
98 units
Official record key
device-enforcement:Z-2784-2017

Official wording

Reason: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Code information: Lot Numbers: 1620000123, 1622800056, 1623500052m 1627700083, 1631200074, 1700300040, 1701000065, 1703700061, 1706500063, 1708000068, 1717000057

Distribution pattern: Nationwide

Derived failure modes

  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    sterile barrier breach