Device product

Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Z-0003-2015

September 02, 2014

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 69258
Status
Terminated
Classification
Class II
Quantity
1
Official record key
device-enforcement:Z-0003-2015

Official wording

Reason: Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images

Code information: Siemens AXIOM Luminos dRF Max system material #s: 10762471 with serial number: 5037.

Distribution pattern: Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images