Device product
Baxter Amia Automated PD systems are used in the treatment of adult renal failure patients to perform automated peritoneal dialysis.
Z-0932-2015
Product summary
- Event
- Event 69586
- Status
- Terminated
- Classification
- Class II
- Quantity
- 55 units
- Official record key
device-enforcement:Z-0932-2015
Official wording
Reason: System error 01779 is produced when the battery cannot be charged due to specific voltage differences between the battery and the charger.
Code information: Product Code: 5C9310
Distribution pattern: Nationwide Distribution including GA, MA, NC, NY, PA, and WA.
Derived failure modes
-
Unknown
System error 01779 is produced when the battery cannot be charged due to specific voltage differences between the battery and the charger.