Skip to content
Recall Observatory FDA recall evidence

Device product

DIASTAT(R) ANA (anti-Nuclear antibody) EURO DIAGNOSTICA, The DIASTAT(R) anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl-70, Jo-1, dsDNA, histone and centromere antigens.

Z-2627-2017

May 22, 2017

Class III

Product summary

Firm
Euro Diagnostica AB
Event
Event 77426
Status
Terminated
Classification
Class III
Quantity
501 units
Official record key
device-enforcement:Z-2627-2017

Official wording

Reason: FANA200 kit lot TS 3577 was re-worked due to shortened shelf-life. The alert sticker on the kit box of the lot informed customers that the shelf-life had been shortened and referenced a letter sent to customers for more information, where the new expiration date was stated. The original kit box label remained on the kit box stating the original expiration date, the packaging did not display the correct expiration date.

Code information: TS 3577

Distribution pattern: MN, TX, NE, PA, Spain, Egypt, Pakistan, Nepal, Canada

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    FANA200 kit lot TS 3577 was re-worked due to shortened shelf-life. The alert sticker on the kit box of the lot informed customers that the shelf-life had been shortened and referenced a letter sent to customers for more information, where the new expiration date was stated. The original kit box label remained on the kit box stating the original expiration date, the packaging did not display the correct expiration date.