PKG, MIXTER FORCEPS, P/N 0250080249. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Z-0637-2015

November 17, 2014

Class II

Product summary

Firm
Stryker Endoscopy
Event
Event 69838
Status
Terminated
Classification
Class II
Quantity
180,573 units total
Official record key
device-enforcement:Z-0637-2015

Official wording

Reason: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code information: 140618 740488 842873 846273 847450 941861 942066 943413 945219 945447 1040727 1041476 1042473 1042696 1042712 1141555 1142624 1142809 1143341 1144351 1144399 1145081 1145410 1145980 1146351 1146683 1242474 1243804 1244258 1343013 1343693 1344289 1345375 1345521 1346227 1440406 171399 646475 743736 943413

Distribution pattern: Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.