Device product
Syngo RT Therapist: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer
Z-1013-2015
Product summary
- Event
- Event 70200
- Status
- Terminated
- Classification
- Class II
- Quantity
- 3
- Official record key
device-enforcement:Z-1013-2015
Official wording
Reason: Combination of CTVision with syngo RT Therapist / syngo RT Oncologist 4.3.SP1 automatic registration in Adaptive Targeting might result in wrong offset calculations. Applying this offset can lead to patient mistreatment. Cone Beam imaging is not affected by this problem.
Code information: model # 08162815, serial #s 10010, 10046, 10586
Distribution pattern: UT, WI, NY
Derived failure modes
-
Unknown
Combination of CTVision with syngo RT Therapist / syngo RT Oncologist 4.3.SP1 automatic registration in Adaptive Targeting might result in wrong offset calculations. Applying this offset can lead to patient mistreatment. Cone Beam imaging is not affected by this problem.