Device product
BEQ-TOP 49101 Childrens Large ECC Pack, Custom Tubing Kit, Catalog No. 701067069R01
Z-1717-2018
Product summary
- Event
- Event 79873
- Status
- Terminated
- Classification
- Class II
- Quantity
- 4
- Official record key
device-enforcement:Z-1717-2018
Official wording
Reason: The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
Code information: Lot 3000059724
Distribution pattern: Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.
Derived failure modes
-
Unknown
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.