Device product

ADVIA Centaur¿ DHEA-SO4

Z-1688-2018

January 02, 2018

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 79505
Status
Terminated
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-1688-2018

Official wording

Reason: The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere

Code information: ADVIA Centaur¿ DHEA-SO4; radioimmunoassay, dehydroepiandrosterone (free and sulfate); SMN 10282227 ADVIA Centaur DHEA-S Kit Lot # 00119053 14041053 80449053 90520053 29750054 30423054 39155054 50533054 63267055 68936055 79273055 86427055 06402057 07109057 14187057 37499057

Distribution pattern: Nationwide and Canada, Mexico, and Thailand

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere