Device product

Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

Z-0939-2013

December 13, 2012

Class II

Product summary

Firm
Draeger Medical, Inc.
Event
Event 64447
Status
Terminated
Classification
Class II
Quantity
260
Official record key
device-enforcement:Z-0939-2013

Official wording

Reason: The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.

Code information: Catalog Number MS20783 (Infinity MCable-Dual Hemo)

Distribution pattern: Distributed in the states of CA, IL, and MA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.