CelluTome Epidermal Harvesting System Harvester (5.0 CM) STERILE EO Rx Only Manufactured for: KCI USA, Inc. Part Number CT-H50 or CT-H25 The CelluTome Epidermal Harvesting System is intended to reproducibly cut a thin skin graft for autologous skin grafting.

Z-0933-2015

November 19, 2014

Class II

Product summary

Firm
KCI USA, INC.
Event
Event 69961
Status
Terminated
Classification
Class II
Quantity
1,139/5 Packs
Official record key
device-enforcement:Z-0933-2015

Official wording

Reason: KCI has received reports that, in a small number of cases, the CelluTome Harvester (a component of the CelluTome Epidermal Harvesting System), blades have the potential to drift during shipping and could prematurely protrude into openings of harvester top plate.

Code information: CT-H25 lot numbers: 3249980000, 3249990000, 3250010000, 2537872, 2556866; CT-H50 lot numbers: 3249660000, 3249840000, 3250050000, 2498886, 2510874, 2537879, 2545871

Distribution pattern: Worldwide Distribution - USA (Nationwide) and Internationally to Australia, Austria, Canada, France, Germany, Ireland, Italy, Isreal, Netherlands, South Africa, Spain, Sweden, Switerland, Turkey, UAE, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    KCI has received reports that, in a small number of cases, the CelluTome Harvester (a component of the CelluTome Epidermal Harvesting System), blades have the potential to drift during shipping and could prematurely protrude into openings of harvester top plate.