Skip to content
Recall Observatory FDA recall evidence

Device product

Neptune¿ 2 Rover - ULTRA ~120V ~ 60 Hz 12A REF 0702-001-000, Rx, Caution, consult accompanying documents Consult instructions for use, Max Vacuum; 21.0 in-Hg, 530 mm-Hg, The Neptune Waste Management System consists of a mobile rover unit used to suction and collect fluid waste, small debris and electrocautery smoke from a surgical site. Mobility allows the rover unit to be relocated to a waste disposal area where the rovers collection canisters can be emptied, via the docking station. Intended to be used in the Operating Room, Surgical Centers and Doctors Offices to collect and dispose of surgical fluid waste as well as collect smoke from electrocautery or laser devices.

Z-0454-2013

May 24, 2011

Class II

Product summary

Firm
Stryker Instruments Div. of Stryker Corporation
Event
Event 63634
Status
Terminated
Classification
Class II
Quantity
5865
Official record key
device-enforcement:Z-0454-2013

Official wording

Reason: Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.

Code information: Part 0702-001-000 (120v) Lot 0733700353 through 1036100423

Distribution pattern: Worldwide Distribution-USA (nationwide) and Pacific countries, and the countries of EEMEA, Canada, Latin America, and Japan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.