Device product
nordicICE 2.3.14 Image processing software package used by trained professionals, including physicians and medical technicians.
Z-2044-2018
Product summary
- Event
- Event 78463
- Status
- Terminated
- Classification
- Class II
- Quantity
- 97 licenses
- Official record key
device-enforcement:Z-2044-2018
Official wording
Reason: An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Code information: Version 2.3.14
Distribution pattern: Worldwide Distribution -- USA, Australia, Denmark, China, France, Japan, Malaysia, Netherlands, USA, Poland, Sweden, Spain, Russia, India, Norway, Singapore, South Korea, United Kingdom, and Slovenia.
Derived failure modes
-
Unknown
An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.