Device product
Dimension¿ Sirolimus (SIRO) Flex¿ reagent cartridge
Z-1670-2018
Product summary
- Event
- Event 79505
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-1670-2018
Official wording
Reason: Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
Code information: SIRO, sirolimus test system, DF306, SMN# 10464331 Lot codes: BB7336 FA8037 FA8081 BB8167 FB8177 FB8219
Distribution pattern: Nationwide and Canada, Mexico, and Thailand
Derived failure modes
-
Unknown
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.