Device product
KING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD¿ Patient Size: >70kg Product Usage: Laryngeal airway
Z-0424-2013
Product summary
- Event
- Event 63580
- Status
- Terminated
- Classification
- Class II
- Quantity
- 263 cases of product. (2,630 devices)
- Official record key
device-enforcement:Z-0424-2013
Official wording
Reason: The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
Code information: REF: LAD-705, Lot #K120413
Distribution pattern: US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
Derived failure modes
-
Unknown
The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO