RayStation Radiation Therapy Treatment Planning System; -- RayStation 3.5, RayStation 4.0, RayStation 4.5 and RayStation 4.7. For RayStation 4.7, the issue applies also to machines with fixed jaws, regardless of MLC/jaw position. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Z-1310-2015

March 04, 2015

Class II

Product summary

Firm
RAYSEARCH LABORATORIES AB
Event
Event 70686
Status
Terminated
Classification
Class II
Quantity
Domestic: 245 units
Official record key
device-enforcement:Z-1310-2015

Official wording

Reason: An issue with photon dose calculation for DMLC (Dynamic MLC) plans for machines where the MLC is positioned above the jaws, e.g. some Elekta linacs. The magnitude of the error depends on the beam model output factor corrections and on the individual DMLC plan characteristics.

Code information: Build numbers 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14 and 4.7.0.15

Distribution pattern: Nationwide Distribution.