Skip to content
Recall Observatory FDA recall evidence

Device product

Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems

Z-3211-2018

August 31, 2018

Class II

Product summary

Firm
Accuray Incorporated
Event
Event 80933
Status
Terminated
Classification
Class II
Quantity
259 units
Official record key
device-enforcement:Z-3211-2018

Official wording

Reason: the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System

Code information: All units with the Ashland Brand distributed between August 1, 2017 and July 31, 2018

Distribution pattern: Worldwide Distribution: United States (nationwide), and countries of: United Kingdom, Germany, France, Finland, Netherlands, Portugal, Hong Kong, Poland, Australia, United Arab Emirates, Thailand, and Japan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System