Skip to content
Recall Observatory FDA recall evidence

Device product

AccuChek Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System.

Z-1733-2015

April 22, 2015

Class II

Product summary

Firm
Roche Diabetes Care, Inc.
Event
Event 71235
Status
Terminated
Classification
Class II
Quantity
35,795
Official record key
device-enforcement:Z-1733-2015

Official wording

Reason: Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u

Code information: All Serial Numbers Models: 1. 05458544001, Accu-Chek Combo Kit mg US/English version. 2. 05504686001, Accu-Chek Combo Kit mg US/Spanish version. 3. 05391512001, Accu-Chek Spirit Combo Replacement Pump US/EN. 4. 05391750001, Accu-Chek Spirit Combo Demo Pump  US. 5. 04759702001, Accu-Chek Spirit Pump Kit US/CA Back Up. 6. 04540492001, Accu-Chek Spirit Pump Kit US/CA. 7. 04759729001, Accu-Chek Spirit Replacement Pump.

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u