PROBEAT WITH DISCRETE SPOT SCANNING SYSTEM Product Usage: Hitachi PROBEAT with DSSS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Z-1838-2015

April 21, 2015

Class II

Product summary

Firm
Hitachi America, Ltd., Power Systems Division
Event
Event 71056
Status
Terminated
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-1838-2015

Official wording

Reason: At a PROBEAT III site installed in Japan, the operator of the system found that the radiation was delivered inconsistent with the treatment plan and stopped the irradiation. The irradiation control system may be potentially affected when rebooting a part of the control system.

Code information: No serial number or lot code. Each system is designed for a specific facility therefore no codes are needed for traceability

Distribution pattern: Distribution in US (Texas) and Japan

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    At a PROBEAT III site installed in Japan, the operator of the system found that the radiation was delivered inconsistent with the treatment plan and stopped the irradiation. The irradiation control system may be potentially affected when rebooting a part of the control system.