Device product
KimVent* Closed Suction System for Adults, 14 F, DSE (Product Code 2210); KimVent* Closed Suction System for Adults, 12 F, DSE (Product 21603); KimVent* Closed Suction System for Adults, 14 F, DSE (Product Code 22103); KimVent* Closed Suction System for Adults, 14 F, DSE (Product Code 22714183) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
Z-2506-2015
Product summary
- Event
- Event 71966
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-2506-2015
Official wording
Reason: The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
Code information: Product Code 2210 - M5054T630, M5082T625, M5089T611, M5117T627; Product Code 21603 - M5089T618, M5103T610; Product Code 22103 - M5089T617; Product Code 22714183 M5082T318
Distribution pattern: Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada
Derived failure modes
-
Unknown
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.