Skip to content
Recall Observatory FDA recall evidence

Device product

ARTHROSCOPY SURGICAL PACK 4/C Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .

Z-2699-2015

July 10, 2015

Class II

Product summary

Firm
Customed, Inc
Event
Event 71721
Status
Terminated
Classification
Class II
Quantity
536
Official record key
device-enforcement:Z-2699-2015

Official wording

Reason: Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.

Code information: Code:900740A Lots: 140915497 exp. 1/31/16 141216360 exp. 12/31/15 141216449 exp. 1/31/16 141216449 exp. 1/31/16 150317655 exp. 3/31/16 150317656 exp. 3/31/16 150318093 exp. 3/31/16 150418420 exp. 4/30/16 150518838 exp. 5/31/16 150619188 exp. 6/30/16 150619405 exp. 6/30/16

Distribution pattern: Distributed Only in Puerto Rico.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    inadequate validation/qualification of sterilization processes
  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    inadequate validation/qualification of sterilization