Device product
Bigliani/Flatow¿ The Complete Solution Fukuda Retractors Large Nonsterile 4305-19-40. Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.
Z-1620-2013
Product summary
- Firm
- Zimmer, Inc.
- Event
- Event 65301
- Status
- Terminated
- Classification
- Class II
- Quantity
- 722 units (includes 2010 market withdrawal)
- Official record key
device-enforcement:Z-1620-2013
Official wording
Reason: This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.
Code information: Item Number: 00-4305-019-40 with Lot Numbers 61588143 and 61612276. 2010 Market Withdrawal: Item #:430501940 Lot #'s: 60768814, 60806551, 60893241, 60946820, 60999665, 61031668, 61110808, 61170678, 61205673, 61244194, 61392456, 61430127
Distribution pattern: Worldwide Distribution-USA (nationwide) including the states of NY, MI, MN, OH, IN, IL, VA, MD, PA, WA, KY, CT, RI, NH, TX, MT, and CA and the countries of Chile, China, South Korea, and Australia.
Derived failure modes
-
Unknown
This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.