Device product

VITROS Chemistry Products CDM PROM (Calibration Data Module) No. 0168, Catalog Number/REF 199 9077, used in conjunction with: VITROS Chemistry Products CRSC DT Slides, Generation (GEN) 82, REF 6802721, and VITROS DT Calibrator Kit, Product Code 1957924, and tested on the VITROS DT60/DT60 II Chemistry System; IVD --- Ortho-Clinical Diagnostics, Inc.

Z-1490-2014

January 29, 2014

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 67509
Status
Terminated
Classification
Class II
Quantity
US: 284 units, Ex-US: 4772 units
Official record key
device-enforcement:Z-1490-2014

Official wording

Reason: Customers were unable to calibrate, or use previously calibrated, VITROS DT CRSC slides (REF/Product Code 6802721), GEN 82, on the VITROS DT II System, when using Calibration Data Module (CDM) Rev. 168 (REF/Product Code 199 9077), as CDM 168 does not contain calibration parameters for DT CRSC.

Code information: Calibration Data Module Rev. 0168, REF 199 9077 used in conjuction with of VITROS Chemistry Products CRSC DT Slides, Generation (GEN) 82, Product Code 6802721.

Distribution pattern: Worldwide Distribution - US Distribution including the states of AR, FL, IA, IL, KY, MI, MN, ND, NY and VA, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Customers were unable to calibrate, or use previously calibrated, VITROS DT CRSC slides (REF/Product Code 6802721), GEN 82, on the VITROS DT II System, when using Calibration Data Module (CDM) Rev. 168 (REF/Product Code 199 9077), as CDM 168 does not contain calibration parameters for DT CRSC.