RayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System, Planning, Radiation Therapy Treatment. RayStation is a software system designed for treatment planning and analysis of radiation therapy.

Z-1419-2013

February 15, 2013

Class II

Product summary

Firm
RAYSEARCH LABORATORIES AB
Event
Event 65136
Status
Terminated
Classification
Class II
Quantity
Domestic: 70 units; Foreign: 27 units
Official record key
device-enforcement:Z-1419-2013

Official wording

Reason: Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.

Code information: Software version and build numbers 2.0.0.15, 2.5.1.89, and 3.0.0.251 (to determine if the version used is affected, bring up the About box in the RayStation application and check the build number reported0.

Distribution pattern: Worldwide Distribution - USA including Connecticut, Florida, Massachusetts, Missouri, New York, Ohio, and foreign distribution to(Canada, Italy, Australia, Germany, Netherlands, and Switzerland.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Under some circumstances, imported CT, MR and PET images are offset from their true positions by one pixel in one or two directions. Users must be aware of this information to avoid incorrect dose calculations during treatment planning.