Device product

Flow-i Anesthesia System Product Usage: The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment by healthcare professionals trained in inhalation anesthesia administration.

Z-0804-2016

December 10, 2015

Class I

Product summary

Firm
Maquet Cardiovascular Us Sales, Llc
Event
Event 73137
Status
Terminated
Classification
Class I
Quantity
7 units within US
Official record key
device-enforcement:Z-0804-2016

Official wording

Reason: Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues. Only the Model C20 is affected in the US.

Code information: Model numbers: C20, C30 and C40 (Only Model # C20 for US) Part numbers: 66 77 200 (US Only), 66 77 300 and 66 77 400 Serial numbers (2743, 2852, 2854, 2855, 2856, 2858

Distribution pattern: US Nationwide in the states of CA, NY, and OR

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues. Only the Model C20 is affected in the US.