Device product
arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
Z-0631-2016
Product summary
- Event
- Event 72860
- Status
- Terminated
- Classification
- Class I
- Quantity
- 11 boxes (25 foil packaged test strips per box)
- Official record key
device-enforcement:Z-0631-2016
Official wording
Reason: One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
Code information: Lot Number PN5C26
Distribution pattern: Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
Derived failure modes
-
Unknown
One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.