Device product
Minstrel (with scale) Product Usage: is a mobile passive hoist. It is intended for lifting and transporting adult residents on horizontal floors in hospitals, nursing homes or other health care facilitates. The Minstrel device aids in the lifting, transfer and lowering to safe position, of a patient. This is accomplished by placing a sling under the patient, and attaching the sling to a spreader bar that is connected to a lifting arm jig via a connection to the spreader bar that allows a swiveling up-down movement. Using battery power, an actuator then powers the lifting arm upwards so that the patient in the sling is lifted from, typically, a bed or chair.
Z-2693-2014
Product summary
- Event
- Event 68594
- Status
- Terminated
- Classification
- Class II
- Quantity
- 18 Lifts
- Official record key
device-enforcement:Z-2693-2014
Official wording
Reason: ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp
Code information: Model Numbers: HMA0002 and HMB002-US; Serial Numbers: MPLAT0409W1177 ; MPLAT0410W1213 ; MPLAT0410W1228 ; MPLAT0410W1229 ; MPLAT0410W1231 ; MPLAT0410W1232 ; MPLAT0412W1290 ; MPLAT0412W1296 ; MPLAT0412W1297 ; MPLAT0502W1403 ; MPLAT0502W1404 ; MPLAT0503W1485 ; MPLAT0503W1486 ; MPLAT0504W1518 ; MPLAT0504W1519 ; MPLAT0504W1520 ; MPLAT0504W1531
Distribution pattern: Worldwide Distribution - US Nationwide in the state of MA and the countries of Australia, Germany, and Italy.
Derived failure modes
-
Unknown
ArjoHuntleigh received three reports where the bolt under the scale unscrewed from the scale. There is a remote risk concerning the bolt unscrewing from the scale. When the scale and bolt parts are not correctly serviced, or when they receive a sharp blow during or outside of use, it is possible the bolt holding the spreader bar to the scale could unscrew during subsequent use. As a result, the sp