Skip to content
Recall Observatory FDA recall evidence

Device product

MRIdian¿ ViewRay¿ Radiation Therapy System, ViewRay¿ Is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Z-0115-2016

August 27, 2015

Class II

Product summary

Firm
Viewray Incorporated
Event
Event 72166
Status
Terminated
Classification
Class II
Quantity
4 units
Official record key
device-enforcement:Z-0115-2016

Official wording

Reason: ViewRay discovered that in the event that an encoder breaks or fails on the Patient Handling System (PHS, or couch), when attempting to restart the system, the couch could move unexpectedly.

Code information: Catalog/Part Number: 10000; Serial Numbers: 100, 101, 102, & 105

Distribution pattern: US Distribution to the states of : CA, MO and WI., and Internationally to Korea.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0