Centricity Cardio Imaging 5.0 is a software only Picture Archiving and Communication System (PACS). Product Usage: It will be sold as a software only device to operate on general purpose computing hardware. Centricity Cardio Imaging receives medical images and other information from various data sources. Information can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Centricity Cardio Imaging is intended to assist trained professionals in the viewing, analysis and diagnostic interpretation of images and other information for the diagnosis and treatment of cardiac and vascular disease. These trained professionals include but are not limited to physicians, cardiologists, radiologists, nurses, medical technicians and assistants. Centricity Cardio Imaging is not intended for use in the patient vicinity.

Z-0034-2014

October 02, 2012

Class II

Product summary

Firm
GE Healthcare It
Event
Event 64787
Status
Terminated
Classification
Class II
Quantity
1 unit
Official record key
device-enforcement:Z-0034-2014

Official wording

Reason: GE Healthcare is aware of a potential safety risk where an image labeling situation exists in the Image Viewer software of the Centricity Cardio Imaging product, version 5.0 SP1. Multiple studies may be displayed side by side by selecting "Compare Study" from the Study List, or by opening prior studies in the Patient Folder. In these cases, the Viewer differentiates the studies wityh a label (Curr

Code information: Software Version: 5.0

Distribution pattern: USA Nationwide Distribution in the state of OK

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software of the Centricity Cardio Imaging product, version 5.0 SP1. Multiple studies may