Skip to content
Recall Observatory FDA recall evidence

Device product

ProScreen 10 Panel Dip Card w/Adult, Item No. PSDA-10MO, PSDA-10MMO300, PSDA-10PPX, PSDA-10BUP, PSD-12BUP, PSD-10M, PSCupA-6MB-W, PSCup-10MMO300, PSD-5AB, PSD-10CMO3, PSCupA-10MO, PSCupA-12M-W I-DXA-1127-023, PSCupA-12TBU.

Z-0519-2016

November 16, 2015

Class III

Product summary

Firm
Ameditech Inc
Event
Event 72651
Status
Terminated
Classification
Class III
Quantity
1,989 kits
Official record key
device-enforcement:Z-0519-2016

Official wording

Reason: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Code information: 144902 150474 151686 151687 152043 152225 152312 152402 152952 153306 153563 153663 153664 153665 153680 153682 153703 153704 153759 153760 153763 153852 153853

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.