Skip to content
Recall Observatory FDA recall evidence

Device product

NxStage System One S Cycler -High Permeability Hemodialysis System Model no. NX1000-4

Z-0337-2016

October 29, 2015

Class II

Product summary

Firm
NxStage Medical, Inc.
Event
Event 72554
Status
Terminated
Classification
Class II
Quantity
7 units
Official record key
device-enforcement:Z-0337-2016

Official wording

Reason: Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

Code information: Software version 4.9 and 4.10

Distribution pattern: Worldwide Distribution.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software error