Device product

Ergon 2 Articulating Equipment Boom Monitor Brackets mounting point for display interfaces utilized in clinical applications for viewing purposes and to provide an articulating point of movement on the ceiling mounted radial arm suspension system of the Ergon 2 equipment pendant / booms

Z-0655-2016

December 02, 2015

Class II

Product summary

Firm
Skytron, Div. The KMW Group, Inc
Event
Event 72897
Status
Terminated
Classification
Class II
Quantity
286
Official record key
device-enforcement:Z-0655-2016

Official wording

Reason: The hardware that holds the VESA plate and the display interface / monitor to the bracket may fail causing the display interface to fall from the bracket. Over time and with use the threaded rod hardware material can weaken and detach from the bracket. A display interface falling from its mounting location may result in injury to the patient or provider.

Code information: Model Number(s): H8-010-50-FC2-LH and H8-010-50-FC2-RH. VESA plates distributed from 09/26/08 to 04/30/13.

Distribution pattern: Worldwide Distribution - US including WI, TX, FL, NC, MD, OH, AZ, CA, UT, MT, OK, HI, CO, WY, MO, NJ, IL, NY, MA, IA, DE, PA, MI, IN, and WV and Internationally to Canada, and Japan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The hardware that holds the VESA plate and the display interface / monitor to the bracket may fail causing the display interface to fall from the bracket. Over time and with use the threaded rod hardware material can weaken and detach from the bracket. A display interface falling from its mounting location may result in injury to the patient or provider.