Skip to content
Recall Observatory FDA recall evidence

Device product

PKG, GRASPER FORCEPS, NARROW JAW, P/N 0250080287. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Z-0657-2015

November 17, 2014

Class II

Product summary

Firm
Stryker Endoscopy
Event
Event 69838
Status
Terminated
Classification
Class II
Quantity
180,573 units
Official record key
device-enforcement:Z-0657-2015

Official wording

Reason: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code information: 848400 941060 944297 945752 1040798 1042479 1042670 1043881 1046197 1142169 1242130 1341574 1341992 1344135 1344687 1344715 1344876 1346231

Distribution pattern: Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods