Device product
00597001333 Femur CR PRC SURF HDN FEM CO-NID CML Rx, Sterile; 00597001334 Femur CR PRC SURF HDN FEM CO-NID CMR Rx, Sterile; 00597001433 Femur CR PRC SURF HDN FEM CO-NID DML Rx, Sterile; 00597001434 Femur CR PRC SURF HDN FEM CO-NID DMR Rx, Sterile; 00597001533 Femur CR PRC SURF HDN FEM CO-NID EML Rx, Sterile; 00597001534 Femur CR PRC SURF HDN FEM CO-NID EMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Z-2010-2013
Product summary
- Firm
- Zimmer, Inc.
- Event
- Event 65413
- Status
- Terminated
- Classification
- Class II
- Quantity
- 192,355 all devices
- Official record key
device-enforcement:Z-2010-2013
Official wording
Reason: Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Code information: all codes
Distribution pattern: Worldwide Distribution-USA (nationwide) and the country of Japan.
Derived failure modes
-
Unknown
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d