Device product
VITROS 5600 Integrated System, System Product Code 6802413. For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Z-0078-2015
Product summary
- Event
- Event 69234
- Status
- Terminated
- Classification
- Class II
- Quantity
- USA: 843; Foreign: 945
- Official record key
device-enforcement:Z-0078-2015
Official wording
Reason: Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied with VITROS Chemistry Products Calibrator Kits 1, 2, 3, 4, 6, or 9, an unexpected assay calibration may occur if assay targets are unassigned (i.e., hidden).
Code information: Software Version 3.0 and below; Serial Numbers J56000110 - J56001995 (serial numbers are sequential; J Numbers are analogous to serial numbers)
Distribution pattern: Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Derived failure modes
-
Device software or design
anomaly with Software