Skip to content
Recall Observatory FDA recall evidence

Device product

PKG, 3MM PEEK MONOPOLAR HANDLE, P/N 0250282045 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Z-0773-2015

November 17, 2014

Class II

Product summary

Firm
Stryker Endoscopy
Event
Event 69838
Status
Terminated
Classification
Class II
Quantity
180,573 units total
Official record key
device-enforcement:Z-0773-2015

Official wording

Reason: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code information: 843744 847017 942768 943108 943436 943515 944197 944825 945711 1040927 1042098 1042468 1042513 1042514 1042957 1043819 1045057 1045626 1046584 1140930 1141656 1142239 1144237 1144342 1145343 1146674 1147037 1242641 1243948 1246163H 1340920H 1341894 1341894H 1344199 1344872 1344990 1345272 1345272H 1346215 1346215H 1346684 1440429 1441830

Distribution pattern: Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods