Device product
Brilliance CT System 6/10/16/16P/40 configuration. No packaging, sold as single unit. Computed tomography scanner intended to provide cross sectional images of the human body and visualization of the internal organs.
Z-2709-2014
Product summary
- Event
- Event 69123
- Status
- Terminated
- Classification
- Class II
- Quantity
- 364
- Official record key
device-enforcement:Z-2709-2014
Official wording
Reason: The system has a gap that can appear between the upper and lower tilt stand covers. The fingers may be squeezed if a person was to place them in this region while the gantry was tilting from an angle to zero tilt angle.
Code information: 728256, 728251, 728246, 728240, and 728235.
Distribution pattern: Worldwide Distribution -- AZ, CO, GA, IA, IL, IN, KS, KY, MD, MI, MN, MS, NC, NY, OH, OK, TX, VT, and WI; and, countries of Singapore, New Zealand, India, Australia, Korea, Japan, Korea, China, India, Taiwan, Australia, Turkey, France, South Africa, Spain, Italy, Austria, Belgium, Switzerland, Israel, Netherlands, England, and Brazil.
Derived failure modes
-
Unknown
The system has a gap that can appear between the upper and lower tilt stand covers. The fingers may be squeezed if a person was to place them in this region while the gantry was tilting from an angle to zero tilt angle.