Skip to content
Recall Observatory FDA recall evidence

Device product

Syngo.Plaza VB 10 A, image processing system (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.

Z-2794-2016

August 02, 2016

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 75125
Status
Terminated
Classification
Class II
Quantity
101 Systems
Official record key
device-enforcement:Z-2794-2016

Official wording

Reason: Siemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values. This suffix stands for on projection and indicates that a shown measurement value may be different from the real value.

Code information: Model Numbers: 10863171, 10863172, 10863173, 10592457 Serial Numbers: 100262,100389,100198,100258,100259,100538,100135,100133,100213,100214,100031,100032,100484,100362,100233,100140,100141,100142,100143,100144,100416,100105,100106,100109,100110,100111,100112,100113,100114,100115,100116,100117,100119,100120,100461,100457,100245,100447,100364,100314,100298,100299,100386,200277,100139,100027,100028,100300,100559,100440,100441,100197,100170,100025,100026,100084,100186,100189,100190,100191,100192,100195,100196,100550,100033,100034,100306,100307,100232,100449,100450,100185,100596,100256,100199,100188,100193,100194,100257,100226,100228,100425,100029,100030,100253,100203,100204,100132,100247,100248,100434,100215,100491,100495,100230,100221,100212,100539,100206

Distribution pattern: Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Siemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values. This suffix stands for on projection and indicates that a shown measurement value may be different from the real value.