Device product

The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or muscles when connected to an extension and/or lead. The Medtronic InterStim II Neurostimulator, Model 3058, and Medtronic InterStim Neurostimulator, Model 3023, are parts of neurostimulation systems for InterStim Therapy. The neurostimulators are multiprogrammable devices that accommodates a lead through which a stimulation program is delivered. The Medtronic Enterra Neurostimulator, Model 3116, is part of an Enterra Therapy system for gastric electrical stimulation (GES). The neurostimulator is a multiprogrammable device that accommodates a lead through which a stimulation program is delivered.

Z-1147-2015

January 09, 2015

Class II

Product summary

Firm
Medtronic Neuromodulation
Event
Event 70271
Status
Terminated
Classification
Class II
Quantity
396,975 (258,805 USA, 138,170 OUS)
Official record key
device-enforcement:Z-1147-2015

Official wording

Reason: Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.

Code information: All lot/serial numbers are affected for all 4 Model numbers: " Model 7425 " Model 3058 " Model 3023 " Model 3116

Distribution pattern: Worldwide Distributon- USA( nationwide including DC and PR) and Internationally to Mexico and Canada

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With cycling enabled, longevity estimate displayed on the Model 8840Programmer may be inaccurate.