Device product

SB Knife -Electrosurgical Knife, Sterile Size: Working Length 2.3M Maximum OD 2.6mm. Product Usage: The SB Knife is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47703W, MD-47704, MD-47706 and MD-47703L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current.

Z-0388-2017

September 09, 2016

Class II

Product summary

Firm
Akita Sumitomo Bakelite Co., Ltd.
Event
Event 75388
Status
Terminated
Classification
Class II
Quantity
15 units
Official record key
device-enforcement:Z-0388-2017

Official wording

Reason: Mislabeled product: Sterilization pouch and carton box labeled as Standard type instead of Jr type-Long

Code information: Lot No.: 60564711

Distribution pattern: US in the states of MA

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    Mislabeled