Device product

Light Sheer Desire Diode Laser System with XC Handpiece Accessory options.

Z-1519-2015

March 25, 2015

Class II

Product summary

Firm
Lumenis Limited
Event
Event 70863
Status
Terminated
Classification
Class II
Quantity
117 units
Official record key
device-enforcement:Z-1519-2015

Official wording

Reason: Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended settings. Operator Manual parameters are lower than indicated for specific hair color and Fitzpatrick skin type resulting in insufficient treatment effect. May result in patient burns and hypopigmentation.

Code information: Lot numbers: XI0017, XI0018, XI0023, XIP0043, 182, 184, 185, XIP0049, 236, 237, 239, 240, 242, 248, 249, 250, 251, 254, 256, 257, XIP0038, XIP0131, XIP0125, 3048, 3052, 3065, 3074, XI0012, XIP0143, XIP0144, XI0001, XIP0072, XI0028, XIP0061, XI0028, XI0036, XIP0074, XIP0077, XIP0082, XIP0087, XIP0132, XIP0127, XIP0124, XIP0129, XI0008, XI0029, XI0037, XI0046, XIP0123, XIP0059, XIP0107, XIP0089, XIP0096, XIP0156, XI0005, XIP0114, XI0014, XI0022, XI0015, XI0021, XI0031, XI0034, XI0024, XI0026, XIP0047, XI0042, XI0039, XIP0044, XIP0025, XIP0035, XIP0041, XIP0030, XIP0051, XIP0057, XIP0113, XIP0067, XIP0053, XIP0083, XIP0058, XIP0055, XIP0060, XIP0066, XIP0163, XIP0071, XIP0063, XIP0120, XIP0088, XIP0081, XIP0064, XIP0050, XIP0101, XIP0062, XIP0095, XIP0121, XIP0094, XIP0108, XIP0105, XIP0122, XIP0119, XIP0104, XIP0097, 104, XIP0065, XIP0164, XIP0159, XIP0152, XIP0166, XI0016, XIP0045, XIP0052, XIP0158, XIP0092, XIP0111, XIP0086, XIP0086, XIP0086, XIP0112, XIP0130, XIP0157, XIP0093, XIP0151, XIP0149, XIP0167, XIP0168, XIP0162, XIP0171, XIP0138, XIP0118

Distribution pattern: Worldwide distribution. US nationwide (to OR, NM, NY, IL, and CO), Switzerland, Canada, Italy, Great Britain, United Arab Emirates, France, Belgium, India, Israel, Japan, Portugal, Czech Republic, Kuwait, Germany, Spain, Georgia, Armenia, Slovenia, New Zealand, Saudi Arabia, and Russian Federation.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended settings. Operator Manual parameters are lower than indicated for specific hair color and Fitzpatrick skin type resulting in insufficient treatment effect. May result in patient burns and hypopigmentation.