Device product
Light Sheer Desire Diode Laser System with XC Handpiece Accessory options.
Z-1519-2015
Product summary
- Firm
- Lumenis Limited
- Event
- Event 70863
- Status
- Terminated
- Classification
- Class II
- Quantity
- 117 units
- Official record key
device-enforcement:Z-1519-2015
Official wording
Reason: Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended settings. Operator Manual parameters are lower than indicated for specific hair color and Fitzpatrick skin type resulting in insufficient treatment effect. May result in patient burns and hypopigmentation.
Code information: Lot numbers: XI0017, XI0018, XI0023, XIP0043, 182, 184, 185, XIP0049, 236, 237, 239, 240, 242, 248, 249, 250, 251, 254, 256, 257, XIP0038, XIP0131, XIP0125, 3048, 3052, 3065, 3074, XI0012, XIP0143, XIP0144, XI0001, XIP0072, XI0028, XIP0061, XI0028, XI0036, XIP0074, XIP0077, XIP0082, XIP0087, XIP0132, XIP0127, XIP0124, XIP0129, XI0008, XI0029, XI0037, XI0046, XIP0123, XIP0059, XIP0107, XIP0089, XIP0096, XIP0156, XI0005, XIP0114, XI0014, XI0022, XI0015, XI0021, XI0031, XI0034, XI0024, XI0026, XIP0047, XI0042, XI0039, XIP0044, XIP0025, XIP0035, XIP0041, XIP0030, XIP0051, XIP0057, XIP0113, XIP0067, XIP0053, XIP0083, XIP0058, XIP0055, XIP0060, XIP0066, XIP0163, XIP0071, XIP0063, XIP0120, XIP0088, XIP0081, XIP0064, XIP0050, XIP0101, XIP0062, XIP0095, XIP0121, XIP0094, XIP0108, XIP0105, XIP0122, XIP0119, XIP0104, XIP0097, 104, XIP0065, XIP0164, XIP0159, XIP0152, XIP0166, XI0016, XIP0045, XIP0052, XIP0158, XIP0092, XIP0111, XIP0086, XIP0086, XIP0086, XIP0112, XIP0130, XIP0157, XIP0093, XIP0151, XIP0149, XIP0167, XIP0168, XIP0162, XIP0171, XIP0138, XIP0118
Distribution pattern: Worldwide distribution. US nationwide (to OR, NM, NY, IL, and CO), Switzerland, Canada, Italy, Great Britain, United Arab Emirates, France, Belgium, India, Israel, Japan, Portugal, Czech Republic, Kuwait, Germany, Spain, Georgia, Armenia, Slovenia, New Zealand, Saudi Arabia, and Russian Federation.
Derived failure modes
-
Unknown
Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended settings. Operator Manual parameters are lower than indicated for specific hair color and Fitzpatrick skin type resulting in insufficient treatment effect. May result in patient burns and hypopigmentation.