Skip to content
Recall Observatory FDA recall evidence

Device product

Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273.

Z-1848-2015

June 15, 2015

Class II

Product summary

Firm
Pega Medical Inc.
Event
Event 71560
Status
Terminated
Classification
Class II
Quantity
4 units
Official record key
device-enforcement:Z-1848-2015

Official wording

Reason: SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation.

Code information: Part Number: SCF-MLD265, LOT#: 130703-01. Part Number: SCF-MLD273, LOT#: 130703-02.

Distribution pattern: US Distribution to the states of : FL and TN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation.