Device product

Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.

Z-1088-2016

January 29, 2016

Class II

Product summary

Firm
Medtronic Navigation, Inc.
Event
Event 73303
Status
Terminated
Classification
Class II
Quantity
1 unit
Official record key
device-enforcement:Z-1088-2016

Official wording

Reason: Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality

Code information: Serial Number: C0936

Distribution pattern: Worldwide Distribution to Austria only.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality