Device product
GE Healthcare Definium 5000 / 5220493 X-Ray System
Z-0703-2013
Product summary
- Event
- Event 64134
- Status
- Terminated
- Classification
- Class II
- Quantity
- 312 installed in US
- Official record key
device-enforcement:Z-0703-2013
Official wording
Reason: When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The exposure can still be made as commanded, however it may not be properly indicated.
Code information: Definium Model 5000/5220493
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The exposure can still be made as commanded, however it may not be properly indicated.