Device product
system, x-ray, angiographic, model # 10094135, 10094137, 10094139, 10094141, 10280959 Artis zee/ zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Z-1299-2016
Product summary
- Event
- Event 73548
- Status
- Terminated
- Classification
- Class II
- Quantity
- 552 units
- Official record key
device-enforcement:Z-1299-2016
Official wording
Reason: In Artis zeego systems, angulations in the vicinity of the C-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. In Artis systems with A100Plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the X-ray tube) can result in the failure of a module in the high-voltage generator.
Code information: Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959
Distribution pattern: US Nationwide Distribution including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and PUERTO RICO.
Derived failure modes
-
Unknown
In Artis zeego systems, angulations in the vicinity of the C-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. In Artis systems with A100Plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the X-ray tube) can result in the failure of a module in the high-voltage generator.