Radiation Therapy Treatment Planning System, Model 5.0
Z-0079-2017
Product summary
- Event
- Event 74750
- Status
- Terminated
- Classification
- Class II
- Quantity
- 491 units (187 domestically & 304 internationally)
- Official record key
device-enforcement:Z-0079-2017
Official wording
Reason: For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.
Code information: 5.0.0.37, 5.0.1.11 and 5.0.2.35
Distribution pattern: AZ, CA, FL, MT, NC NY, TX, OH & WA
Derived failure modes
-
Unknown
For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.