Device product

Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used for viewing medical images. eFilm Workstation with Modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). eFilm Workstation with Modules can be used to communicate, process and display medical images. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on medical images to aid in preoperative planning. eFilm Workstation with Modules can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record. Typical users of eFilm Workstation with Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians

Z-0707-2017

October 20, 2016

Class II

Product summary

Firm
Merge Healthcare, Inc.
Event
Event 75270
Status
Terminated
Classification
Class II
Quantity
2163
Official record key
device-enforcement:Z-0707-2017

Official wording

Reason: A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after calibration is performed which should be present if the measurement is presented based on Pixel Spacing.

Code information: EFilm Workstation and eFilm Lite versions 2.1, 2.1.2, 3.0, 3.1, 3.3.5, 3.4, 4.0, 4.0.1, 4.0.2, 4.0.3 and 4.1.

Distribution pattern: Worldwide Distribution - US: Nationwide including DC, GU, PR and Internationally to Canada, Argentina, Armenia, Australia, Austria, Belgium, Brazil, Chile, China, Colombia, Congo, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Guatemala, hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mauritania, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, and Yemen.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after calibration is performed which should be present if the measurement is presented based on Pixel Spacing.