Skip to content
Recall Observatory FDA recall evidence

Device product

SIEMENS IMMULITE/IMMULITE 1000 Systems Rubella IgM (RUM), REF/Catalog Number LKRM1 (OUS), Siemens Material Number (SMN) 10381282 (OUS), Unique Device Identification Number (UDI) 00630414964577; IVD --- For in vitro diagnostic use with the IMMULITE/IMMULITE 1000/IMMULITE 2000/IMMULITE 2000 XPi Systems Analyzers-for the qualitative detection of IgM antibodies to Rubella virus in human serum or plasma (EDTA and heparinized), as an aid in the presumptive diagnosis of an acute or recent rubella infection, particularly in women of childbearing age.

Z-0081-2017

August 24, 2016

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 75037
Status
Terminated
Classification
Class II
Quantity
Foreign: 359 Kits
Official record key
device-enforcement:Z-0081-2017

Official wording

Reason: Siemens received multiple customer complaints indicating an increase in false positive and indeterminate patient sample results and increased imprecision for several lots of Rubella IgM assay on the IMMMULITE 2000/IMMULITE 2000 XPi platforms.

Code information: Kit lots 0333 and 0334 (OUS)

Distribution pattern: Worldwide Distribution n- US (Nationwide) and Foreign distribution to the following countries: Argentina, Bangladesh, Brazil, Bulgaria, China, Colombia, Costa Rica, Cyprus, Egypt, France, Germany, Greece, Guatemala, India, Indonesia, Israel, Italy, Kazakhstan, Macedonia, Mexico, Paraguay, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Africa, Syria, United Arab Emirates, and Uruguay.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Siemens received multiple customer complaints indicating an increase in false positive and indeterminate patient sample results and increased imprecision for several lots of Rubella IgM assay on the IMMMULITE 2000/IMMULITE 2000 XPi platforms.