Device product

Chord-X Chordal Sizer is designed to assist in determining the proper length of chordal replacement prostheses to repair the mitral valve and help the surgeon select a preconfigured prosthesis, or configure the correct size using straight suture material, during mitral valve repair.

Z-1057-2017

December 13, 2016

Class II

Product summary

Firm
On-X Life Technologies, Inc.
Event
Event 75951
Status
Terminated
Classification
Class II
Quantity
45 units
Official record key
device-enforcement:Z-1057-2017

Official wording

Reason: Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plastic particulate was found within sealed primary packages, 3) Loose hair was discovered within four sealed primary packages and 4) Assembly error, wherein Left and Right Housing components were acceptably press-fit together, was discovered within one sealed primary package. One of the Left Housing press-fit pins was not pressed into the associated hole of the Right Housing, resulting in a small gap in the housing assembly.

Code information: 72-00-47420

Distribution pattern: CA, WA, NE, DE, OH, ME, TX

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    particulate
  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    Assembly error