Skip to content
Recall Observatory FDA recall evidence

Device product

***REF 606-01-05***OPTEFORM Allograft Paste 5 cc, bone void filler of bony defects in dental and orthopedic applications.***STERILE Using Radiation. ***Rx Only***Distributor: EXACTECH 2320 NW 66th Court, Gainesville, FL 32653 USA***Manufacturer: RTI Biologicals, Inc. 11621 Research Circle, Alachua, FL 32615 USA.

Z-0939-2014

December 06, 2013

Class II

Product summary

Firm
RTI Surgical, Inc.
Event
Event 67246
Status
Terminated
Classification
Class II
Quantity
14 units
Official record key
device-enforcement:Z-0939-2014

Official wording

Reason: RTI Surgical, Inc. of Alachua, FL is recalling their OPTEFORM Allograft Paste, 5cc as it may contain a 10 cc fluid dispenser. The 10cc fluid dispenser does not include the correct fluid fill line reference for reconstitution of 5cc past allografts.

Code information: Serial Numbers: 8504395, 8504396, 8504397, 8504398, 8504399, 8504450, 8504452, 8504453, 8504454, 8504455, 8504456, 8504457, 8504458 and 8504459.

Distribution pattern: Distributed to Florida.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    RTI Surgical, Inc. of Alachua, FL is recalling their OPTEFORM Allograft Paste, 5cc as it may contain a 10 cc fluid dispenser. The 10cc fluid dispenser does not include the correct fluid fill line reference for reconstitution of 5cc past allografts.